Chapter 4 Remote Operation Mode

The DB9

interface connector

on the rear panel of the power supply can be transferred to RS-232

interface, the following information will tell you how to use the computer to control the output of the
power supply.

4.1 IT-E131 Communication cable

The DB9 interface connector on the rear panel of power supply is TTL voltage level; you can use

the communication cable (IT-E131) to connect the DB9 interface connector of the power supply
and the RS-232 interface connector of computer for the communication.

Computer side TTL? RS232 Cable (IT-E131) PS side

PC

Load

IT-E131 communication cable

COMPUTER

INSTRUMENT

RX

TX

IT-E131 ISOLATED

COMMUNICATION CABLE

TTL(5V)

RS232

ISOLATION

859666668889942311

IT

COMPUTER

INSTRUMENT

RX

TX

IT-E131 ISOLATED

COMMUNICATION CABLE

TTL(5V)

RS232

ISOLATION

859666668889942311

IT

Power

supply

PC



4.2 Communication setting

Before using the remote operation mode, please make sure that the baud rate and communication

address in power supply are the same as in the computer software, otherwise, the communication will
fail, you can change the baud rate and communication address from the front panel or from computer.

1. Address: the range is from 0 to 254, default setting is 0
2. Baud rate: 4800,9600,19200 and 38400 are selectable, default setting is 9600
3. Data bit: 8 bit
4. Stop bit: 1
5. Parity: None

Start

Bit

8 Data Bits

Stop

Bit

PARITY = NONE

Note:

It will not work if you connect the DB9 interface connector of the power

supply to the RS232 interface connector of computer directly by a standard
RS232 cable. Please use IT-E131 to connect them.