Save the calibration data to eeprom ( 2dh), Setting calibration information (2eh), Reading calibration information (2fh) – B&K 1787B User Manual

Page 25

Advertising
background image

25





13. Sending the actual output current to calibration program (2CH)

1

st

byte

Start bit (AAH)

2

nd

byte

Address(0~0XFE)

3

rd

byte

Command(2CH)

4

th

byte

The lower byte of the present current value

5

th

byte

The higher byte of the present current value

6

th

to 25

th

byte

System reserve

26

th

byte

Check sum



14. Save the calibration data to EEPROM( 2DH)

1

st

byte

Start bit(AAH)

2

nd

byte

Address (0~0XFE)

3

rd

byte

Command(2DH)

4

th

to 25

th

byte

System reserve

26

th

byte

Check sum






15. Setting calibration information (2EH)

1

st

byte

Start bit (AAH)

2

nd

byte

Address (0~0XFE)

3

rd

byte

Command(2EH)

4

th

to 23

rd

byte

Calibration information(ASIC code)

24

th

byte

System reserve

25

th

byte

System reserve

26

th

byte

Check sum



16. Reading calibration information (2FH)

1

st

byte

Start bit (AAH)

2

nd

byte

Address (0~0XFE)

3

rd

byte

Command( 2FH)

4

th

to 23

rd

byte

Calibration information(ASCII code)

24

th

byte

System reserve

25

th

byte

System reserve

26

th

byte

Check sum



17. Reading product’s model, series number and version information (31H)

1

st

byte

Start bit (AAH)

Note: To calibrate the 2 points of the current value sequentially.