8 . entering the calibration mode ( 27h) – B&K 1787B User Manual

Page 23

Advertising
background image

23

11

th

byte

To set the low byte of current value

12

th

byte

To set the high byte of current value

13

th

byte

Byte 0 of the maximum voltage value

14

th

byte

Byte 1 of the maximum voltage value

15

th

byte

Byte 2 of the maximum voltage value

16

th

byte

Byte 3 of the maximum voltage value

17

th

byte

Byte 0 of output voltage value

18

th

byte

Byte 1 of output voltage value

19

th

byte

Byte 2 of output voltage value

20

th

byte

Byte 3 of output voltage value

21

st

to 25

th

byte

System reserve

26

th

byte

Check sum














8. Entering the calibration mode( 27H)

1

st

byte

Start bit(AAH)

2

nd

byte

Address(0~0XFE)

3

rd

byte

Command(27H)

4

th

byte

Calibration protection state

5

th

byte

Calibration password(0X28H)

6

th

byte

Calibration password(0X01H)

7

th

to 25

th

byte

System reserve

26

th

byte

Check sum












9. Reading the calibration state (28H)

Note:

1. We use 4 bytes to represent the maximum voltage value as follows:

Byte 3

Byte 2 Byte1

Byte
0

2. We use 1 byte to represent power supply’s state. Each bit is defined as

follows:

From higher bit to lower bit

7

6

5

4

3

2

1

0

0 bit: The output state, 0 is OFF, 1 is ON.
1 bit: Over heat protection, 0 is normal, 1 is abnormal.
2? 3 bit: The output mode, 1 is CV mode, 2 is CC mode,3 is Unreg mode.
4? 5? 6 bit: The fan speed, 0 is stop, 5 is the maximum fan speed.

7 bit: Operation state, 0 is front panel operation mode, 1 is remote control

mode.

3. The frame format is the same as above

Note:

We use a byte to represent calibration protection state, each bit is defined as

follows:

from higher bit to lower bit

7

6

5

4

3

2

1

0

0 bit: Protection state, 0 is to disable protection, 1 is to enable the protection.