HF scientific MicroTOL Online Turbidimeter old series- 0-1000 NTU User Manual

Page 19

Advertising
background image

MICRO TOL (12/04)
Rev. 2.7

13







2. Remove the flow through unit.
3. Insert the requested 1000 NTU standard. Index the standard to the lowest value on the

upper display.

4. Press

the

↵ button to accept the calibration.

5. The lower display will count down the progress of the calibration step.
6. The lower display will now change to show alternating 10 and

↵, requesting the 10.0

NTU standard.










7. If the alternating 10 and

is not displayed, push the t or

uuntil you get this

display.

8. Insert the requested 10.0 NTU standard. Index the standard to the lowest value on the

upper display.

9. Press

the

button to accept the calibration.

10. The lower display will count down the progress of the calibration step.

11. The lower display will now change to show 02 and

, requesting the 0.02 NTU

standard.










12. Insert the requested 0.02 NTU standard. Index the standard to the lowest value on the

upper display.

13. Press the

↵ button to accept the calibration.

14. The lower display will count down the progress of the calibration step.
15. The instrument will show alternating END and

.

16. Press the

↵ button to accept the calibration.

17. The instrument will display SAVE on the lower display briefly and then the Mode

will return to AUTO.