HF scientific MicroTOL Online Turbidimeter old series- 0-1000 NTU User Manual

Page 29

Advertising
background image

MICRO TOL (12/04)
Rev. 2.7

23

Select the correct baud rate (1200, 2400, 4800, or 9600) for operation of the I/O port by
pressing either the

t or u button to change the displayed baud rate. When you have

selected the proper baud rate press the

button to continue on to 7.7 Setting the Security

Access Option. The communication parameters are 8 bits, no parity and 1 stop bit.

Note: The information printed out in the timed printout has the format: x.xx NTU dd mm

yyyy hh:mm. The resolution of the turbidity level output is the same as the
resolution of the turbidity level on the display.

Bi-Directional Communication: The instrument is equipped with the ability for bi-
directional communication. This means that a remote unit can prompt the instrument to
send out the current turbidity levels via the RS-485. If the timed print option was turned
OFF, the next prompt will be for the address of the Micro TOL.

Here you can select the desired instrument address (1-255) using the

t or u buttons.

Once you are satisfied you can press the

↵ button to select. You will then be requested to

select the BAUD.

Select the correct baud rate (1200, 2400, 4800, or 9600) for operation of the I/O port by
pressing the

t or u buttons to change the displayed baud rate. When you have selected

the proper baud rate press the

button to continue on to 7.7 Setting the Security Access

Option. The communication parameters are 8 bits, no parity and 1 stop bit.