HF scientific MicroTOL Online Turbidimeter old series- 0-1000 NTU User Manual

Page 28

Advertising
background image

MICRO TOL (12/04)
Rev. 2.7

22

Select the turbidity level you wish to assign to the UPLM using the

t and u buttons.

Once you have set the desired level, press the

↵ button to accept it. The next prompt will

be for the access code. Refer to section 7.7 Setting the Security Access Option.

7.6 Configuring the RS–485 I/O Port (if equipped)
If your instrument is equipped with the option, the following menus will be displayed
after the date and time.
Automatic/Timed Printouts: After pressing the

↵ button, PRNT will be displayed. The

Upper row of the display will indicate whether this option is turned ON or OFF. Turn
ON to select timed printouts.

Next, select the printing interval. This is the interval at which the instrument will print
out the instantaneous turbidity level, time, and date. Your choices are 1 HOUR, 4
HOUR
, 8 HOUR, 15 MIN & 30 MIN. Use the

t or u button to change and the ↵

button to select. You will then be requested to select the BAUD.

Note: The timing for the interval starts at midnight.