8.3 Diagnostic Chart

Symptom Cause Cure

Lower display shows MA

4-20 mA loop open

Check wiring. See sections
3.3.4 and 7.5.3

Lower display shows DESC

Desiccant pouch bad

Change desiccant pouch. See
section 9.2

Lower display shows CAL

Calibration failure

Refer to sections 5 and 6.1

Lower display shows LAMP

Lamp failed

Replace lamp. Refer to section
9.3

Lower display shows FLOW

Sample flow has stopped

Restore flow. Contact HF about
factory installed option

Lower display shows FAIL

Major system fault

Refer to section 8.1 & 8.2

Readings are higher than
expected

(1) Bubbles in solution









(2) Condensate or leaky

cuvette


(3) Flow through cuvette

dirty

Instrument out of calibration

(1)Ensure that the drain vent is
open and is not obstructed.
See section 3.2.2.
(2)Apply backpressure. See

section 3.2 and figure 4

(3) For sever cases of bubbles a

stilling chamber is
available. Call HF scientific.

Part# 20106

Check flow through cuvette for

condensate or leaks.


Clean cuvette. See section 9.1


Recalibrate. Refer to section 5

Readings are erratic

(1) Bubbles in solution
(2) Debris in flow through

(1) See above
(2) Clean debris from cuvette

Readings are lower than expected Instrument out of calibration

Recalibrate. Refer to section 5

Upper display shows O-r

Sample Over-Range

Check sample. Must be over
1000 NTU.

8.4 Technical and Customer Assistance
If for any reason you need assistance regarding this instrument please do not hesitate to
contact either the HF scientific, inc. Technical Service Department or the HF scientific,
inc. Customer Service Department:

HF scientific, inc.

3170 Metro Parkway

Fort Myers, Florida 33916-7597

Phone: (239) 337-2116

Fax: (239) 332-7643

Email: [email protected]

Advertising