Procedure – LaMotte SMART3 Colorimeter User Manual

Page 192

Advertising
background image

PROCEDURE

1. Press and hold

until colorimeter turns on.

2. Press

ENTER

to select TESTING MENU.

3. Select

ALL TESTS (or another sequence containing 050 Iodine) from

TESTING MENU.

4. Scroll to and select 050 Iodine from menu.
5. Rinse a clean tube (0290) with sample water. Fill tube to the 10 mL line with

sample.

6. Insert tube into chamber, close lid and select SCAN BLANK.
7. Remove tube from colorimeter. Add one *DPD #1 Tablet Instrument Grade

(6903A). Cap and shake 10 seconds. Invert slowly 5 times. Solution will turn

pink if iodine is present. Wait 15 seconds. Mix.

8. Insert tube into chamber, close lid and select SCAN SAMPLE. Record result.
9. Press

to turn colorimeter off or press

EXIT

to exit to a previous menu or

make another menu selection.

IODINE, DPD, Tablet

SMART3 Test Procedures 11.10