OZONE

INDIGO METHOD • CODE 3651-SC

QUANTITY

CONTENTS

CODE

15 mL

Chlorine Inhibitor

3990-E

250 mL

*Ozone Buffer

*3991-K

30 mL

Indigo Blue Stock Solution

3989-G

1

Sampling Apparatus

0681

1

Pipet, transfer, 1.0 mL

2-2170

1

Pipet, transfer, 5 mL

0329

1

Pump, 10 mL

30527

1

Bottle, HR Reagent, amber glass

3989-MT

1

Graduated Cylinder, 50 mL, glass

0418

*WARNING: Reagents marked with an * are considered to be potential health

hazards. To view or print a Material Safety Data Sheet (MSDS) for these reagents

go to www.lamotte.com. To obtain a printed copy, contact LaMotte by e-mail,

phone or fax.

Ozone is sometimes used in place of, or in conjunction with, chlorine or other

halogens for disinfection of pool, spa, or drinking waters. Recently, large aquatic

facilities have begun using ozone as a disinfectant in many artifi cial habitats.

APPLICATION:

Drinking, pool and aquatic waters.

RANGE:

0.00–0.40 ppm Ozone, Low Range

0.00–3.00 ppm Ozone, High Range

MDL:

0.02 ppm Ozone, Low Range

0.05 ppm Ozone, High Range

METHOD:

Ozone rapidly and stoichiometrically decolorizes Indigo

Trisulfonate under acidic conditions.

SAMPLE HANDLING

& PRESERVATION:

Ozone is extremely unstable in aqueous solutions. Test

must be performed immediately and the sample must

not be agitated.

INTERFERENCES:

Manganese at any level interferes.

SMART3 Test Procedures 11.11

OZONE, Indigo