Procedure – LaMotte SMART3 Colorimeter User Manual

Page 291

Advertising
background image

PROCEDURE

A. PREPARATION OF DILUTE ZINC INDICATOR SOLUTION

1. Use a pipet (0352) to add exactly 5.0 mL of *Zinc Indicator Solution (6314)

to a 10 mL graduated cylinder (0416). The bottom of the curved surface (the

meniscus) of liquid should be at 5.0 mL mark. Pour this into the bottle labeled

“Dilute Zinc Indicator Solution” (6314-MT).

2. Use unrinsed graduated cylinder to add 10.0 mL and then 7.8 mL (total of 17.8

mL) of *Methyl Alcohol (6319) to bottle labeled “Dilute Zinc Indicator Solution”

(6314-MT). Cap and mix ingredients in this bottle. Do not leave this bottle

uncapped.

B. DETERMINATION OF ZINC

1. Press and hold

until colorimeter turns on.

2. Press

ENTER

to select TESTING MENU.

3. Select

ALL TESTS (or another sequence containing 099 Zinc LR) from

TESTING MENU.

4. Scroll to and select 099 Zinc LR from menu.
5. Rinse a clean tube (0290) with sample water. Fill to the 10 mL line with sample.
6. Insert tube into chamber, close lid and select SCAN BLANK. (See Note)
7. Remove tube from colorimeter. Use 0.1 g spoon (0699) to add one measure

of Sodium Ascorbate Powder (6316). Use 0.5 g spoon (0698) to add one

measure of *Zinc Buffer Powder (6315). Cap and shake vigorously for 1

minute. Some undissolved buffer may remain in the bottom of the tube.

8. Add 3 drops of *Sodium Cyanide, 10% (6565). Cap and mix.
9. Use the 1 mL pipet assembly to add 1 mL of “Dilute Zinc Indicator Solution”.

Cap and mix.

10. Use a second plain pipet (0352) to add 4 drops of *Formaldehyde Solution,

37% (5128). Cap and mix by inverting 15 times.

11. Insert tube into chamber, close lid and select SCAN SAMPLE. Record result.
12. Press

to turn colorimeter off or press

EXIT

to exit to a previous menu or

make another menu selection.

NOTE: For best possible results, a reagent blank should be determined to account

for any contribution to the test result by the reagent system. To determine the

reagent blank, follow the above test procedure to scan a distilled or deionized

water blank. Then follow the above procedure to perform the test on a distilled or

deionized water sample. This test result is the reagent blank. Subtract the reagent

blank from all subsequent test results of unknown samples. It is necessary to

determine the reagent blank only when a new lot number of reagents is obtained.

SMART3 Test Procedures 11.11

ZINC, Low Range

Test P

rocedures

Advertising