Appendix – LaMotte SMART3 Colorimeter User Manual

Page 292

Advertising
background image

APPENDIX

Ammonia in water occurs in two forms: toxic unionized ammonia (NH

3

) and the

relatively non-toxic ionized form, ammonium ion (NH

4

+

). This test method measures

both forms as ammonia-nitrogen (NH

3+

–N) to give the total ammonia-nitrogen

concentration in water. The actual proportion of each compound depends on

temperature, salinity, and pH. A greater concentration of unionized ammonia is

present when the pH value and salinity increase.

1. Consult the table below to fi nd the percentage that corresponds to the

temperature, pH, and salinity of the sample.

2. To express the test result as ppm Unionized Ammonia Nitrogen (NH

3

–N),

multiply the total ammonia-nitrogen test result by the percentage from the

table.

3. To express the test result as ppm Ammonia Nitrogen (NH

3+

–N), subtract the

unionized ammonia-nitrogen determined in step 2 from the total ammonia-

nitrogen.

10°C

15°C

20°C

25°C

pH

FW1

SW2

FW

SW

FW

SW

FW

SW

7.0

0.19

0.27

0.40

0.55

7.1

0.23

0.34

0.50

0.70

7.2

0.29

0.43

0.63

0.88

7.3

0.37

0.54

0.79

1.10

7.4

0.47

0.68

0.99

1.38

7.5

0.59

0.459

0.85

0.665

1.24

0.963

1.73

1.39

7.6

0.74

0.577

1.07

0.836

1.56

1.21

2.17

1.75

7.7

0.92

0.726

1.35

1.05

1.96

1.52

2.72

2.19

7.8

1.16

0.912

1.69

1.32

2.45

1.90

3.39

2.74

7.9

1.46

1.15

2.12

1.66

3.06

2.39

4.24

3.43

8.0

1.83

1.44

2.65

2.07

3.83

2.98

5.28

4.28

8.1

2.29

1.80

3.32

2.60

4.77

3.73

6.55

5.32

8.2

2.86

2.26

4.14

3.25

5.94

4.65

8.11

6.61

8.3

3.58

2.83

5.16

4.06

7.36

5.78

10.00

8.18

8.4

4.46

3.54

6.41

5.05

9.09

7.17

12.27

10.10

8.5

5.55

4.41

7.98

6.28

11.18

8.87

14.97

12.40

1

Freshwater data from Trussel (1972).

2

Seawater values from Bower and Bidwell (1978).

Salinity for Seawater values = 34% at an ionic strength of 0.701m.

ZINC, Low Range

SMART3 Test Procedures 11.10

Test P

rocedures

Advertising