Procedure – LaMotte SMART3 Colorimeter User Manual
Page 256
PROCEDURE
1. Press and hold
until colorimeter turns on.
2. Press
ENTER
to select TESTING MENU.
3. Select
ALL TESTS (or another sequence containing 079 Phosphate HR) from
TESTING MENU.
4. Scroll to and select 079 Phosphate HR from menu.
5. Rinse a clean tube (0290) with sample water. Fill to the 10 mL line with sample.
6. Insert tube into chamber, close lid and select SCAN BLANK.
7. Remove tube from colorimeter. Use the 1.0 mL pipet (0354) to add 2.0 mL of
*VM Phosphate Reagent (4410). Cap and mix. Wait 5 minutes for full color
development.
8. After 5 minute waiting period, mix, insert tube into chamber, close lid and select
SCAN SAMPLE. Record result.
9. Press
to turn colorimeter off or press
EXIT
to exit to a previous menu or
make another menu selection.
PHOSPHATE, High Range
SMART3 Test Procedures 11.10
Test P
rocedures