Procedure – LaMotte SMART3 Colorimeter User Manual

Page 256

Advertising
background image

PROCEDURE

1. Press and hold

until colorimeter turns on.

2. Press

ENTER

to select TESTING MENU.

3. Select

ALL TESTS (or another sequence containing 079 Phosphate HR) from

TESTING MENU.

4. Scroll to and select 079 Phosphate HR from menu.
5. Rinse a clean tube (0290) with sample water. Fill to the 10 mL line with sample.
6. Insert tube into chamber, close lid and select SCAN BLANK.
7. Remove tube from colorimeter. Use the 1.0 mL pipet (0354) to add 2.0 mL of

*VM Phosphate Reagent (4410). Cap and mix. Wait 5 minutes for full color

development.

8. After 5 minute waiting period, mix, insert tube into chamber, close lid and select

SCAN SAMPLE. Record result.

9. Press

to turn colorimeter off or press

EXIT

to exit to a previous menu or

make another menu selection.

PHOSPHATE, High Range

SMART3 Test Procedures 11.10

Test P

rocedures

Advertising