24

2B. Waste and Water Bottles

1. Empty the waste container, and check the diluent

water bottle for sufficient volume.

2C. Sample Tube Installation

1. Add sample to 15 ml or 50 ml conical tubes.

2. Place sample tube lids on the conical tubes.

3. Lift the sipper tube away from the instrument.

4. Insert the conical tube for each position.

2D. Fraction Tube Installation

1. Insert bottom of tube into the fraction tube

holder indent.

2. Tilt the top of the tube into the top holder and

push into place.

Numbering of Collection Tubes

Sample 1

Sample 2

Flow-through

1A

2A

Wash – 1

1B

2B

Wash – 2

1C

2C

Purified protein

1D

2D

Correct installation of
conical tube in the
sample 1 position.

Water volume
required for
selected run

Correct installation of fraction collection tubes in
positions 1A through 1D. Setup shown is for one
sample, as indicated on the display screen
diagram above.

2 1

Sample
sipper tube

Sample lid
on conical
tube