The Bio-Rad methods function provides two sample loading flow rate options:

Standard — recommended for normal- to high-expression proteins (<100 kD)

Low — may improve the yield of low-expression proteins, high MW proteins (>100 kD)
or samples containing detergents

The Bio-Rad methods function provides two cartridge wash options:

Standard — recommended for most purification conditions

Extended — may improve the purity of product for some sample types

Parameters for each type of method are provided in Table 6.1.

Table 6.1. Sample loading Flow rates and wash volumes by cartridge size for each method
type option.

Method Type

Cartridge

Sample Loading

Wash-1

Size, ml

Flow Rate, ml/min

Column

Volume (CV)

Standard

Low

Standard

Extended

IMAC, Native

1

2

0.5

6

10

5

10

2.5

6

10

IMAC, Denaturing

1

2

0.5

6

10

5

10

2.5

6

10

GST, 1

1

0.5

12

15

Protein A and G

5

5

2.5

12

15

eXact

1

1

14

5

5

10

To select the flow rate(s) and wash time(s) for your sample(s):

1. Touch the arrow button to select either the “Standard” or “Low” flow rate for each

sample.

2. Touch the arrow button to select either the “Standard” or “Extended” wash option for

each sample.

Note: If a method will be performed with two samples, make flow rate and wash time
selections separately for each sample.

When the method options are selected, touch the Next button to proceed to the Enter Run
& Sample Information screen. A View Method button will be available in the lower toolbar to
view the individual method steps for the sample(s).

6.3.2 Selecting Sample Incubation Time

Each eXact Bio-Rad method is available with an option to select from three different incubation
times, 0.5 hrs, 2.0 hrs, or 15 hrs (Figure 6.7).