Cooper Instruments & Systems M5-1000 High Capacity Digital Force Gage User Manual

Page 22

Advertising
background image

CF

183

19

32-1113

REV

2

1010

Optionally exercise the load cell shaft several times (at full scale, if possible), then press ENTER.

8. The display will appear as follows:

Apply a weight equal to the full scale of the instrument, then press ENTER.

9. After displaying “PLEASE WAIT…” the display will appear as follows:

22

Remove the load applied in Step 8, leave the fixtures in place, then press ZERO.

10. The display will appear as follows:

Use the UP and DOWN keys to adjust the load value as required. The load values default to even increments,
as indicated by the previously entered number of data points (even increments are recommended for best
results). For example, if a 50 lbF capacity gauge is calibrated, and 5 data points were selected, the load values
will default to 10, 20, 30, 40, and 50 lb. Apply the calibration load. Then press ENTER.

Repeat the above step for the number of data points selected.

11. After all the compression calibration points have been completed, the display will appear as follows:

Press ENTER.

12. The display will appear as follows:

Advertising
This manual is related to the following products: