Reverse the orientation of the load cell shaft by rotating the gauge 180 degrees. Press DIRECTION to invert the
display. Then attach weight fixtures. The following screens will step through the same procedure as with the
compression direction. Proceed in the same manner.

13. At the completion of the tension calibration, the display will appear as follows:

To save the calibration information, select “SAVE & EXIT”. To exit without saving the data select “EXIT W/O
SAVING”.

14. Any errors are reported by the following screens:

Displayed at the start of calibration if a disallowed unit is selected.

Ensure that the load is not swinging, oscillating, or vibrating in any manner. Then try again.

The calibration weight does not match the set value.

The entered calibration point is too close to the previous point.

24

11 PASSWORDS

Two separate passwords may be set to control access to the Calibration section and to the menu and other
keys. To access the passwords setup screen, select Passwords from the menu. The display will appear as
follows: