VII. Battery Replacement / Installation

1. To access the batteries, remove the isolation chamber by depressing the tabs on

the sides of the unit and slide off upwardly.

2. Pull back on the finger tab on the top of the battery door. The battery door should

hinge open to expose the batteries.

3. The batteries may be removed by pressing firmly on the positive end (opposite the

compressed spring) and lifting up.

4. Be certain to dispose of used batteries in a manner consistent with local

environmental and institutional policy for lithium or alkaline battery disposal.

5. To install new batteries, place the negative end on the spring. Press down firmly to

compress spring and tilt battery into place.

6. To replace the battery door, insert the tab on the bottom of the door into the slot

on the bottom of the unit to create a hinge. Swing the top half of the door
upwards pressing firmly on the top of the door until the small tab on the main unit
“clicks” into the catch on the battery door.

7. Reinstall isolation chamber with probe attached.

VIII. Troubleshooting

1. If device fails to function properly after washing & reconnecting probe, rinse all

contacts in de-ionized water and blow dry.

2. Icons indicate all other possible errors and remedies.
3. Device Error Codes:

E01 - System error during synchronization

E02 - System error during patient thermistor calibration

E03 - System error during heater thermistor calibration

E04 - System timing error

E05 - Heater error

E06 - LCD communication error

E07 - Shut off error

P01 - Probe configuration (or no probe connected) error

P02 - Direct Mode patient thermistor unstable or out of range

P03 - Direct Mode heater thermistor unstable or out of range

P04 - Predict Mode patient thermistor unstable or out of range

P05 - Predict Mode heater thermistor unstable or out of range

P06 - Unable to pre-heat probe tip

4. No user-serviceable parts inside.
5. Refer to a factory or qualified Biomedical Technician for service.