IX. Specifications

Medical Electrical Equipment

Filac 3000 EZ/EZA Electronic Thermometer

Classified with respect to electrical shock, fire, and mechanical hazards only in accordance with UL60601-1

Classified with respect to electrical shock, fire, and mechanical hazards only in accordance with CAN/CSA C22.2 No. 60601-1-08

Dimensions (Approximate, without base):

in.- 3.2 W x 6.0 H x 2.1 D ; mm - 84 W x 152 H x 56 D

Dimensions (Approximate, with base):

in.- 4.1 W x 7.5 H x 3.1 D ; mm - 105 W x 190 H x 79 D

Weight (Approximate, without base):

13.5 Ounces; 385 Grams

Temperature Measurement Range:

30°C to 43°C; 86°F to 109°F

Ambient Operating Temperature Range:

10°C to 40°C; 50°F to 104°F; at 10% - 95% RH, non-condensing.

Average prediction times
(after insertion into measurement site):

Oral (EZ model):

3-5 seconds (non-fever temps),

8-10 seconds (fever temps)

Oral (EZA model):

6-10 seconds

Axillary Mode:

8-12 seconds

Rectal Mode:

10-14 seconds

Direct Mode (All Sites):

60-120 seconds

Typical times for switch to Direct Mode:

If no measurement site detected: 60 seconds

(after pulling probe from probe well)

If temperature does not stabilize: 70 seconds

(after measurement site insertion)

Pulse Timer:

60 Second count with a “beep” at 15 seconds, 2 “beeps” at 30 seconds, 1 “beep” at
45 seconds and 2 “beeps” at 60 seconds

Water Bath Accuracy
(35.5°C/95.9°F to 42.0°C/107.6°F):

Direct Mode:

+/-0.1°C - +/- 0.2°F

Standard Prediction Mode**:

+/-0.1°C - +/- 0.2°F

Quick Prediction Mode (Oral Only)**:

+/-0.3°C - +/- 0.5°F

**Greater than 95% of the prediction mode readings will be within the specified accuracy.


Patient Accuracy:

In standard predict mode, thermometer accuracy meets EN 12470-3

A Standard Prediction Mode reading and a Direct Mode reading will differ by

less than ±0.2°C (±0.4°F) on 98% of tested patients

Batteries:

Four “AA” Required. Standard IEC package size. Alkaline – 1.5 Volt

Approximately 6000 temperature readings.

Probe:

Replaceable.

For probe integrity/safety checks, refer to qualified Biomedical Technician or
return to Covidien.

Isolation Chamber:

Removable, replaceable, washable, Polycarbonate-polyester blend.