COVIDIEN Filac™ 3000 EZ/EZA Electronic Thermometer User Manual

Page 12

Advertising
background image

Filac

3000 EZ/EZA Electronic Thermometer

9

Table of Contents

Device Materials:

Thermometer Case:

Flame retardant Polycarbonate/Polyester Blend

Isolation Chamber:

Flame retardant Polycarbonate/Polyester Blend

Probe Handle:

Flame retardant Polycarbonate/Polyester Blend

Probe Shaft:

Flame retardant Polyester

Probe Cable:

Polyurethane jacket with TPE over mold

Tip:

Aluminum

All materials are latex-free.

Type of Protection Against Electrical Shock:

Internally Powered Equipment. 4 AA Batteries.

Degree of Protection Against Electrical Shock: Type BF

Degree of Protection Against Ingress of Fluids: Not Protected-IPX0

Mode of Operation:

Continuous Use. Intended for hand held use.

Degree of Safety of Application in the Presence: Not suitable for use in the presence of flammable anesthetic mixture with air,

oxygen or nitrous oxide.

Transport and Storage:

Not to exceed 30 days

Transport and store this device between temperatures of
-25°C to 55°C (-13°F to131°F).

Relative humidity range should be between 10%-95% RH, non-condensing.

Return unit to the factory if the device is dropped and performance appears degraded, or in the event that the environmental
conditions for transport, storage and operation are exceeded.

The Filac 3000 EZ/EZA Electronic Thermometer is in compliance with the following International Regulatory and

Safety standards:
UL60601-1

CAN/CSA C22.2 No. 60601-1-08

EN60601-1:2005

EN60601-1-2:2001

EN 12470-3:2000

MDD 93/42/EEC

EN980:2008

ASTM E 1112: 2006

Advertising