COVIDIEN Filac™ 3000 EZ/EZA Electronic Thermometer User Manual

Page 16

Advertising
background image

Filac

3000 EZ/EZA Electronic Thermometer

13

Table of Contents

Guidance and manufacturer’s declaration – electromagnetic immunity

The Filac 3000 Electronic Thermometer is intended for use in the electromagnetic environment specified below. The user of the Filac
3000 Electronic Thermometer should assure that it is used in such an environment.
Immunity test

IEC 60601 test level

Compliance level Electromagnetic environment guidance

Electrostatic discharge
(ESD) (EN 61000-4-2 per
EN 60601-1-2: 2001)

± 6 kV contact
± 8 kV air

± 6 kV contact
± 8 kV air

Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30 %.

Electrical fast transient /
burst IEC 61000-4-4

± 2 kV for power supply lines
± 1 kV for input/output lines

not applicable
not applicable

Mains power quality should be that of a typical
commercial or hospital environment.

Surge IEC 61000-4-5

± 1 kV differential mode
± 2 kV common mode

not applicable
not applicable

Mains power quality should be that of a typical
commercial or hospital environment.

Voltage dips, short
interruptions and voltage
variations on power
supply input lines IEC
61000-4-11

< 5 % UT (>95 % dip in UT )
for 0,5 cycle
40 % UT (60 % dip in UT )
for 5 cycles
70 % UT (30 % dip in UT )
for 25 cycles
< 5 % UT (>95 % dip in UT )
for 5 sec

not applicable

Mains power quality should be that of a typical
commercial or hospital environment. If the
user of the Filac 3000 Electronic Thermometer
requires continued operation during power mains
interruptions, it is recommended that the P-STIM be
powered from an uninterruptible power supply or
a battery.

Power frequency (50/60
Hz) magnetic field (EN
61000-4-8 per EN 60601-
1-2: 2001)

3 A/m

3 A/m

Power frequency magnetic fields should be at
levels characteristic of a typical location in a typical
commercial or hospital environment.

Note UT is the a. c. mains voltage prior to application of the test level.

Recommended separation distances between portable and mobile RF communications equipment

and the Filac 3000 Electronic Thermometer

The Filac 3000 Electronic Thermometer is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the Filac 3000 Electronic Thermometer can help prevent electromagnetic interference by
maintaining the minimum distance between portable and mobile RF communications
Rated maximum output power
of transmitter

W

Separation distance according to frequency of transmitter

m

150 kHz to 80 MHz

d= 1,2√P

80 MHz to 800 MHz

d=1,2√P

800 MHz to 2,5 GHz

d = 1,2√P

0,01

0,12

0,12

0,23

0,1

0,38

0,38

0,73

2

1,2

1,2

2,3

10

3,8

3,8

7,3

100

12

12

23

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be
estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
Note 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflections from
structures, objects and people.