ITC Hemochron Response Whole Blood Coagulation System User Manual

Page 7

Advertising
background image

7

SETTING SUPERVISOR OPTIONS
The supervisor can specify default values for heparin units, target ACT, heparin and protamine dose
adjustments, and patient height and weight units. The supervisor can also specify whether the additional

heparin calculation screen is automatically displayed when an ACT result falls below the target ACT. Finally,
the supervisor can erase the entire RxDx database, erase a single case, re-start a closed case, temporarily
deactivate the RxDx module, and restore the factory default settings for RxDx.

Erasing the RxDx Database or a Single Case

1.

Display page 3 of the Supervisor menu:

Note: The supervisor must exit RxDx to display the Supervisor menu.

2.

Press 1 to erase the RxDx database. Press YES to confirm.

3.

Press 2 to erase a case. Enter the PID for the case at the prompt and press YES.

Re-Starting a Closed Case

1.

Display page 3 of the Supervisor menu.

2.

Press 3. Enter the PID for the case at the prompt and press YES.

Temporarily Deactivating the RxDx Module

1.

Display page 3 of the Supervisor menu.

2.

Press 4 to change RxDx On to NO (or vice versa).

Specifying Default Heparin Units

1.

Display page 3 of the Supervisor menu.

2.

Press 5 (RxDx Settings) to display the Customize RxDx screen:

3.

Press 1 to change Hep Units from U/mL to mg/kg (or vice versa).

Specifying the Default Value for Target ACT for HRT, PRT, or PDA-O

1.

Display page 3 of the Supervisor menu.

2.

Press 5 (RxDx Settings) to display the Customize RxDx screen.

3.

Press 2 to display the Target ACT screen:

4.

Select the test (HRT, PRT, or PDA-O) for the target ACT.

5.

Enter the new default value.

6.

Press YES to save the value.