8

Specifying the Default Percentage for Adjusting the Dose
The percentage dose adjustment is the factor by which the calculated dose is increased prior to being
displayed. The factory default settings are 10% for heparin and 10% for protamine. These are the same as
on the RxDx calculator, but different from the HEMOCHRON 8000 system, which uses default settings of
10% for heparin and 30% for protamine.

1.

Display page 3 of the Supervisor menu.

2.

Press 5 (RxDx Settings) to display the Customize RxDx screen.

3.

Press 3 to display the Dose Adjust screen:

4.

Select the dose (heparin or protamine) adjustment to change.

5.

Enter the new default value (from 0 to 30).

6.

Press YES to save the value.

Specifying Default Units for Patient Height and Weight

1.

Display page 3 of the Supervisor menu.

2.

Press 5 (RxDx Settings) to display the Customize RxDx screen.

3.

Press 4 to change Units from m, kg to ft, in, lb (or vice versa).

Specifying Display of the Additional Heparin Calculation Screen

1.

Display page 3 of the Supervisor menu.

2.

Press 5 (RxDx Settings) to display the Customize RxDx screen.

3.

Press 5 to change the display from AUTO to MANUAL (or vice versa).

Restoring RxDx Factory Defaults

1.

Display page 3 of the Supervisor menu.

2.

Press 6. The factory default settings are restored.

Defining a Note
Up to nine customized notes can be defined. Up to two of these notes can be selected and appended to a test

record by the operator when the test is run.

Note: Refer to the HEMOCHRON Response Operator's Manual for instructions to define a note.