Example for transparent transfer of a SPA command

SPA identifier 0 (0x20) command: read current value „R“ (0x52)

Excerpt of operating manual SPA N140 :


This command is for reading the 6-digit current value (bottom line of the display). Data volume is always 6
Byte. A negative value comes with “-“ sign (2Dh) followed by 5 data Bytes. Positive values are returned as 6
data Bytes less sign. Values of less than 6 (5) digits come with preceding zeroes.

Example: Current value = -32,50

SOH Adr Cmd EOT CRC

Transmission

01h 20h 52h 04h 40h

SOH Adr Cmd

Current value = –32,50

EOT CRC

Response

01h 20h 52h 2Dh 30h 33h 32h 35h 30h

04h

54h




More examples in the annex.