0 milstd462d method rs101 testing, Operational cautions, 0 mil-std-462d method rs101 testing – ETS-Lindgren 7606 Coil Antenna/Shielded Coil User Manual

Page 23

Advertising
background image

5.0 MIL-STD-462D Method RS101 Testing

Before connecting any components, follow the
safety information in the ETS-Lindgren

Product Information Bulletin

included with your

shipment.

The normal design application of the Model 7605 Radiating Loop and
Model 7606 Radiating Sensor is the MIL-STD-462D Method RS101 EMI testing.
Additional applications are possible, but only the application to Method RS101 is
discussed here.

Method RS101 requires two distinct activities: test setup calibration and exposure
of the Equipment Under Test (EUT) to specified magnetic flux densities.

Operational Cautions

Large magnetic fields are produced during
testing, particularly at low frequencies. They are
capable of affecting calculators and watches,
and at the RS101 worst-case limits, they exceed
the levels which the Swedish government [10]
has set for video display terminals. Therefore,
take these precautions:

Keep all calculators, watches, rings, and

other metallic objects at least one meter
from the Model 7605 during testing.

Keep your head and torso at least 90 cm

from the Model 7605 during testing below
400 Hz.

The Model 7606 may be damaged by large
currents, so do not connect it to the
signal source, and do not leave it in place
during testing.

MIL-STD-462D Method RS101 Testing

| 23

Advertising
This manual is related to the following products: