Calibration steps – ETS-Lindgren 7606 Coil Antenna/Shielded Coil User Manual

Page 28

Advertising
background image

C

ALIBRATION

S

TEPS

1.

Set the signal source to a frequency of 1 kHz and adjust the output to
provide a magnetic flux density of 110 dB(pT) as determined by the
reading obtained on instrument A and the relationship
B = 160 dB(pT/A). This requires a current of -50 dB(A) or 3.16 mA.

To determine the current, algebraically add the current probe
correction factor in dB(S) to the reading in dB(mV) on instrument A; for
example:

2.

Measure the voltage output in dB(µV) from the Model 7606 on
instrument B.

3.

Verify that the output from the Model 7606 is 42 dB(µV) ±3 dB, and
record this value.

If the output from Model 7606 is not within ±3 dB of 42 dB(µV), verify the source
current is set correctly, and then look for other causes. Some possible causes
are:

Inaccuracy in either or both measuring instruments.

Malfunction of either or both measuring instruments.

Inaccuracy in the current probe.

Loose clamping or other malfunction of the current probe.

Damage to the current probe.

Magnetic-field coupling between Model 7605 and the current probe or
measuring instruments.

Loose fit between Model 7605 and Model 7606, or other incorrect
assembly.

Damage to Model 7605 or Model 7606.

Metallic objects, particularly magnetic materials, in close proximity to
Model 7605 and Model 7606.

28

|