Common Diagnostic Techniques

29

S

IGNAL

D

EMODULATION

To demodulate a signal:

1.

Set the spectrum analyzer for a 0 Hz frequency span and tune to the

signal of interest. This essentially changes the spectrum analyzer into

a tuned receiver and makes the display a frequency filtered

oscilloscope.

2.

Take the video output off the spectrum analyzer and run it to the

oscilloscope. Using the oscilloscope as the display allows greater

flexibility in adjusting the signal amplitude and in triggering.