ETS-Lindgren 7405 E & H Near Field Probe Set User Manual

Page 47

Advertising
background image

www.ets-lindgren.com

Common Diagnostic Techniques

47

The more distant measurement points may lose the signal into the

system noise; a given solution may only redirect the beam. Especially

with narrow beam problems, solutions frequently only shift the beam so

that it radiates in a different direction.

After measurement points are chosen, baseline the unit by measuring

each point with an E-field and an H-field probe. That way, each design

alternative can be implemented and measured over the same set of

points.

2.

The two procedures differ here in how they approach the

measurements that have been taken.

The first method is based upon finding a solution with a large safety

margin. For example, suppose a signal fails the required limit by 3 dB.

Once that signal is found in the lab, it can be measured in the

near-field. The goal is then to reduce in this near-field the 3 dB plus a

safety factor of 6 dB or 10 dB.

This allows a large margin of error due to near-field effects.

Additionally, a solution that passes this must then be confirmed by

far-field measurements.

The second method identifies several solutions which could be

effective. In the previous example where the signal failed by 3 dB, after

pre-screening in the lab, a variety of solutions may be selected and

tested.

A final benefit of pre-screening is that through the inevitable failures, new

information can be discovered. For example, an attempt to reduce an emission

may fail the following reasons:

1.

The diagnosis was wrong.

2.

The technique was inappropriate to the diagnosis.

Advertising