Common Diagnostic Techniques

37

Distance:

A to B =

1.5

2.0

3.0

Propagating Field:

1/R

–3.52 dB

–6.02 dB

–9.54 dB

Reactive Field:

1/R

3

–10.57 dB

–18.06 dB

–28.63 dB

After the source is identified, two or three angles of approach are measured.

A typical situation would record two points at 0.5 meters and 1.5 meters from the

source along two radials from the source. The signal is measured at each point

with a probe which is highly selective of the H-field and another probe which is

highly selective of the E-field. The rate of fall-off is noted for each probe and the

relative amplitude between the probes is noted. In deciding what the relative

amplitude is, the conversion factor of each probe must be taken into account.