Specifications

|

15

5.0 Specifications

5.1 Operational Specifications

Frequency Range:

30 Hz to 50 MHz (maximum usable
bandwidth)

Low Frequency Roll Off:

The low frequency roll off is switch
selectable to be 3 dB down at 22 kHz,
1.9 kHz, or 170 Hz.

High Frequency Roll Off:

Antenna factor is 3 dB down at 30 MHz

Antenna Factor:

See data graphs included with manual. The
calibration graphs were taken in the low
gain setting. Unless otherwise noted, the
low frequency roll off is set at the 170 Hz
3 dB roll off point.

Saturation Level:

• 0.7 V/m narrowband without attenuation

and at the low gain setting

• 22 V/m with 30 dB of internal attenuation

selected

• 63 dBuV/m/MHz broadband saturation

level

Minimum Discernible Signal:

See data graphs included with manual. For
best sensitivity the unit should be in the low
gain setting with all internal attenuation off.

Dynamic Range:

• 115 dB at mid-band (1 MHz)

• 145 dB with the use of the internal

attenuators

Archived 3/18/10