Principles of Operation

|

21

The antenna rod terminal is directly connected to the high

impedance gate of a field effect transistor. Do not lift the

antenna by the rod or touch it before properly grounding out

to the unit case. Accumulated static charge on test personnel

may damage the FET. If the antenna is used in a location

where static discharges are likely to be prevalent, attach a

grounding clip between the chassis and the antenna rod

while the antenna is being set up to prevent damage to the

amplifier circuitry.

The following diagram reflects the sequence of functions as they are connected

on the board.

10 dB (+/- 10%) Input Attenuator

30 dB (+/- 10%) Input Attenuator

1.9 kHz Low Frequency Roll Off

22.0 kHz Low Frequency Roll Off

Moving any switch to the extreme left, toward the center of the unit, activates the

indicated function. Moving the switch to the extreme right, toward the side of the

unit, deactivates that function. The most commonly used configuration will have

the 22 kHz roll off switch on and all other options off.

6.3.1 Factory Settings

The unit comes configured from the factory with the 22.0 kHz LF Roll Off set to

on. All other switches are off.

It is anticipated that the unit will normally be used with one of the roll off switches

on. Due to the abundance of low frequency ambient noise, in most settings using

the full bandwidth of the unit will make it very susceptible to saturation. However,

when extremely low frequency measurements are required, the unit is capable of

providing these measurements.

FR

O

N

T

P

A

N

E

L

ON

22.0 kHz

LF Roll Off

1.9 kHz

LF Roll Off

30 dB

Atten.

10 dB

Atten.

OF

F

FU

S

E

FUSE