Page — 6

HI

n

4456 Manual

Figure 3

n

n

1

Switch and Connectors

connectors. The cables are color-coded-white

for XMIT, yellow for RCV. Identically-colored

dots are located on the electronics housing

adjacent to these connectors. Be sure that

each cable is attached to the proper probe

connector.

When the fiber optic cables are not attached,

always cover the probe connectors with the

protective plastic covers supplied with the

unit, or with similar material. This prevents

dirt or other contaminants from entering the

connector and causing communication

problems.

ARM/OFF

The ARM/OFF switch activates and

deactivates the probe. In the ARM position,

the probe is powered by its internal 3.6 VDC

NiCd battery: in the OFF position, the probe is

inactive. To prolong battery life, set the

ARM/OFF switch to OFF at the end of a test

sequence or when the probe is not in use.