return. If the command does not require the probe to

return any data, the probe simply responds with the

command letter and a carriage return. If an error occurs,

the probe responds with an error code, as detailed in

Appendix A.

Commands

Command

Description

B

Read battery voltage.

Cx

Set baud rate.

x = 1 sets rate to 2400 baud

x = 2 sets rate to 9600 baud

New baud rate does not take effect until the next

power-up.

Dx

Read probe data.

x = 1 enables short form output

x = 2 enables long form output

Lx

Load table data. x = ASCII hex data

Rx

Set range. x = 1, 2, 3, 4 or N (next range)

Sx

Sleep timer. x = number of seconds to wait for

a command before putting probe into the sleep

mode.

Tx

Read Temperature. x = C or F

Ux

Set unit type. x = 1, 2, 3, or N (next unit)

1 = V/m

2 = mW/cm²

3 = [V/m]²