Archived 4/2/10 – ETS-Lindgren HI-4456 Isotropic Electric Field Probe (Archived) User Manual
Page 30
HI
n
4456 Manual
Page — 25
APPENDIX B
HI
n
n
4456 OPERATING PROTOCOLS
Introduction
The information in this appendix assumes that you have
purchased the optional HI
n
4413G Fiber Optic / RS232
Interface, and are capable of communicating directly with
the HI
n
4456 probe via computer. No system readout is
required when using this configuration.
Communication Protocol
Data Type:
RS-232 Serial
Data Mode:
Asynchronous
Word Length:
7 bit
Parity:
Odd
Stop Bits:
1
Data Rate:
9600 baud
Information Transfer Protocol
The HI
n
4456 operates as a Controller Mode device. This
probe only responds to commands from another device;
it transmits no data without first receiving instructions to
do so.
Command Structure
A command to an HI
n
4456 probe consists of 1) a
command letter, followed by 2) possible parameters, 3)
terminated with a carriage return. When it completes the
command, the probe responds with a string consisting of
1) a start character (":"), 2) the command letter, followed
by 3) data, if required, and terminated with 4) a carriage
Archived 4/2/10