Archived 4/2/10 – ETS-Lindgren HI-4456 Isotropic Electric Field Probe (Archived) User Manual

Page 33

Advertising
background image

Page — 28

HI

n

4456 Manual

Probe Output

Command

HI

n

4456 Response

B

Bxx.xx, where xx.xx is the battery voltage.

D1

Dxx.xxuuu, the short form output.

xx.xx is the reading. The position of the decimal

point depends upon the range setting of the

probe.

uuu = units

_V_ = V/m, mW2 = mW/cm², _V2 = [V/m]²

(underscore indicates a space character)

D2

Dxx.xxuuurrrobaaat, the long form output.

xx.xx = the reading, as described for D1.

uuu = units, as describe for D1.

rrr = recorder out value (A 3-digit ASCII number

from 0 to 255).

o = over range indicator ("N" = ok, "O" = over

range).

b = battery status ("N" = safe operating level,

"W" = warning level, "F" = fail level).

aaa = axis enable ("E" = enabled).

t = terminating carriage return.

Rx

Rx, where x is the range.

x = "" returns the range currently in use

x = 1, 2, 3, 4 enables the selected range

x = N sets the probe to the next (higher) range.

Archived 4/2/10