0 calibration, 1 front panel calibration, Calibration – Rice Lake 120 Digital Weight Indicator User Manual

Page 25

Advertising
background image

Calibration

21

4.0

Calibration

The 120 can be calibrated using the front panel, EDP commands, or the Revolution® configuration utility. Each
method consists of the following steps:

Zero calibration

Entering the test weight value

Span calibration

Optional rezero calibration for test weights using hooks or chains.

The following sections describe the calibration procedure for each of the calibration methods.

WZERO

*CAL*

Display and edit

zero calibration

A/D count value

Display and edit

test weight value

WVAL

*CAL*

Display and edit

span calibration

A/D count value

WSPAN

*

CAL*

Press Enter to

remove offset from

zero and span

calibrations

REZERO

VERS

DATE

TIME

PFORMT

SERIAL

CALIBR

CONFIG

FORMAT

PROGRM

10000

Figure 4-1. Calibration (CALIBR) Menu

4.1

Front Panel Calibration

To calibrate the indicator using the front panel, do the
following:

1. Place the indicator in configuration mode

(display reads

CONFIG

) and remove all weight

from the scale platform. If your test weights
require hooks or chains, place the hooks or
chains on the scale for zero calibration.

2. Press

until the display reads

CALIBR

(see

Figure 4-1). Press to go to zero calibration
(

WZERO

).

3. With

WZERO

displayed, press to calibrate

zero. The indicator displays

*CAL*

while

calibration is in progress then the display goes
to (

WVAL)

.

4. With

WVAL

displayed, place test weights on

the scale and press to show the test weight
value. Use the procedure shown in Figure 4-2
to enter the actual test weight, then press to
save the value and go to span calibration
(

WSPAN

).

5. With

WSPAN

displayed, press to calibrate

span. The indicator displays

*CAL*

while

calibration is in progress, then the display
goes to (

REZERO)

.

6. The

REZERO

function is used to remove a

calibration offset when hooks or chains are
used to hang the test weights.

Note

If no other apparatus was used to hang the
test weights during calibration, remove the
test weights and go to Step 7.

If hooks or chains were used during
calibration, remove these and the test
weights from the scale. With all weight
removed, press to rezero the scale. This
f u n c t i o n a d j u s t s t h e z e r o a n d s p a n
calibration values. The indicator displays

*CAL*

while the zero and span calibrations

are adjusted.

7. Press

until the display reads

EXIT Y

, then

press to exit configuration mode.

When editing numeric values, press to allow numeric mode

change entry, then press or to change the digit selected.

Press or to increment or decrement the value of the flashing

selected digit.

Press to save the value entered and return to the level above.

Figure 4-2. Editing Procedure for Numeric Values