0 appendix, 1 error messages, 2 continuous output (stream) format – Rice Lake 120 Digital Weight Indicator User Manual

Page 33: Appendix, Sp ace> <time> < / > <date> <space> <raw_count

Advertising
background image

Appendix

29

7.0

Appendix

7.1

Error Messages

Table 7-1.

120

Error Messages

Error Message

Description

Solution

_ _ _ _ _ _

(bottom LED segments lit)

Display overflow (negative)

Negative weight value too large to be
displayed (< –99999)

_ _ _ _ _ _

(middle LED segments lit)

Overload

Weight value exceeds scale capacity

_ _ _ _ _ _

(top LED segments lit

)

Display overflow (positive)

Positive weight value too large to be
displayed (> 999999)

7.2

Continuous Output (Stream) Format

Figure 7-1 shows the continuous output format sent to the 120 EDP or printer port when the STREAM parameter
(SERIAL menu) is set to either EDP or PRN.

<STX> <POL> <wwwwwww> <UNIT> <G/N> <S> <EXTEND> <TERM>

ASCII 02
(decimal)

Polarity:
<Space> = Positive
<–> = Negative

Weight data: 7 digits, right-justified, with
decimal point, leading zero suppression.
Error =

- - - - -

-

L = pounds
K = kilogram
T = ton
G = grams
O = ounces
<space> = none

G = Gross
N = Net

Status:
<space> = valid
I = Invalid
M = In motion

<CR> <LF>

or <CR>

ASCII 13, 10

(decimal)

<SP

ACE> <TIME> < / > <DATE> <SPACE> <RAW_COUNT>

ASCII 32
(decimal)

ASCII 32
(decimal)

ASCII 47
(decimal)

Time: 8 digits, hh:mm:ss
hh: 00–24
mm: 00–59
ss: 00–59

Date: 2 digits, dd
dd: 01–31

Raw count: 8 digits
00000000–16777216

Figure 7-1. Continuous Output Data Format

Advertising