Rice Lake RL550 Digital Chair Scale Technical and Operating Instructions User Manual

Page 18

Advertising
background image

14

Rice Lake 550 Digital Chair Scale Technical Manual

LOAD

This is the amount of weight
applied during calibration.
Can also be changed in the
calibration menu.

Value (200 Lb)

The display toggles between a numeric value and LOAD.
If you don’t want to change this value, press the BMI key to
move to the next setting. Example: from LOAD to ASTART. If
you want to change the value, use the following steps.

1.

Press the ENTER key to change value starting with
the right most digit.

2.

Use the Up/Down arrow keys to increment/
decrement numbers.

3.

Press the BMI key to move to the left.

4.

Use the Up/Down arrow keys to increment/
decrement numbers.

5.

When done, press the ENTER key to move to the
next parameter (ASTART).

ASTART

Weight process start limit —
Maximum (full capacity)/10.
Determine when the weight
algorithm starts (when the
“----” is displayed), below this
value the scale will show live
weight.

Value (2.0)

The display toggles between a numeric value and ASTART.
If you don’t want to change this value, press the BMI key to
move to the next setting. Example: from ASTART to ARW. If
you want to change the value, use the following steps.

1.

Press the ENTER key to change value starting with
the right most digit.

2.

Use the Up/Down arrow keys to increment/
decrement numbers.

3.

Press the BMI key to move to the left.

4.

Use the Up/Down arrow keys to increment/
decrement numbers.

5.

When done, press the ENTER key to move to the
next parameter (ARW).

ARW

Auto Reweigh — Restarts the
weight algorithm if the weight
changed by more than this
value.

Value (4.0 Lb)

The display toggles between a numeric value and ARW.
If you don’t want to change this value, press the BMI key to
move to the next setting. Example: from ARW to SAL. If you
want to change the value, use the following steps.

1.

Press the ENTER key to change value starting with
the right most digit.

2.

Use the Up/Down arrow keys to increment/
decrement numbers.

3.

Press the BMI key to move to the left.

4.

Use the Up/Down arrow keys to increment/
decrement numbers.

5.

When done, press the ENTER key to move to the
next parameter (SAL).

SAL

Semi Auto Live —This value is
the interval between weight
displays during the algorithm
process.

Value (0.5)

The display toggles between a numeric value and SAL.
If you don’t want to change this value, press the BMI key to
move to the next setting. Example: from SAL to ROUND. If
you want to change the value, use the following steps.

1.

Press the ENTER key to change value starting with
the right most digit.

2.

Use the Up/Down arrow keys to increment/
decrement numbers.

3.

When done, press the ENTER key to move to the
next parameter (ROUND).

Parameter

Description

Choices

Steps

Table 5-1. Configuration Mode Menu (Continued)

Advertising
This manual is related to the following products: