Rice Lake 550 Digital Chair Scale - Troubleshooting and Testing

25

8.0

Troubleshooting and Testing

Refer to the following to check and correct any failure before contacting service personnel.

Table 8-1. Troubleshooting Table for the Rice Lake Scale Line

Symptom

Possible Cause

Corrective Action

Scale does not turn on

Dead battery

Connect the scale to a power source.

Faulty electrical outlet

Use a different electrical outlet.

Bad power supply

Replace optional adapter.

Questionable weight or the
scale does not zero

External object is interfering
with the scale.

Remove the interfering object from the scale.

Display did not show 0.0 before
weighing.

Help the patient off the scale, zero the scale and begin the weighing process
again.

Scale is not placed on a level
floor.

Ensure the scale is level and begin the weighing process again.

Scale is out of calibration.

Check the weight with a known weight value.

The display shows E messages as detailed below.

E06

Identifier - ADC

AD too high

E07

AD too low

E10

Overload

Scale has been overloaded. Remove load from scale.

E4L

BAT

Battery low but still useable

1 bar left on the indicator display.

E4U

Battery low and unstable

no bars left on the indicator display.

E11

CAL

Calibration Error

recalibrate the scale again.

Err 2

Low saturation state (low A/D)

The load cell is not connected properly. Check the cables and mechanical
connections. If the problem persists, replace the set of load cells.

Err 3

High saturation state (high A/D)

See Err 2.

Err 6

Unstable weight. Cannot
calibrate.

Check the load cells’ mechanical surroundings and see that nothing touches
them and that the cables are properly welded.

SAT

Damaged load cell cable

Replace load cell cable.

9.0

Maintenance

The following section provides instructions for maintaining and cleaning the Rice Lake line of scales.

Maintenance operations other than those described in this section should be performed by qualified service

personnel.

9.1

Basic Maintenance

Before the first use of the scale and after periods of non-use, check the scale for proper operation and function. If

the scale does not operate correctly, contact qualified service personnel.
Go through the following steps for basic maintenance.

1. Check the overall appearance of the entire scale for any obvious signs of damage.
2. Inspect the condition of the optional AC adapter for cord cracking or fraying or for broken or bent

prongs.

9.2

Cleaning

Proper care and cleaning is essential to ensure a long life of accurate and effective operation. Before beginning

the cleaning process, disconnect the scale from the AC power source.

1. Clean all external surfaces with a clean, damp cloth or tissue. Mild soap and water solution may be used.

Dry with a clean soft cloth.

2. Do not immerse the scale in cleaning or other liquid solutions.
3. Do not use isopropyl alcohol or other solutions to clean the display surface.