0 calibration, 1 front panel calibration, 2 edp command calibration – Rice Lake SURVIVOR LV Series Livestock Ring Scale User Manual

Page 31: Calibration

Advertising
background image

Calibration

27

4.0

Calibration

The 590 AG Indicator can be calibrated using the front panel, EDP commands, or Revolution

®

. Each method

consists of the following steps:

Zero calibration

Entering the test weight value

Span calibration

Optional rezero calibration for test weights using hooks or chains.

The following sections describe the calibration procedure for each of the calibration methods.

XXXXXXX

XXXXXXX

WZERO

*CAL*

Display and edit

zero calibration

A/D count value

Display and edit

test weight value

WVAL

*CAL*

Display and edit

span calibration
A/D count value

WSPAN

VERS

PROGRM

PFORMT

SERIAL

CALIBR

CONFIG

FORMAT

Press Enter to

remove offset from

zero and span

calibrations

REZERO

*CAL*

Figure 4-1. Calibration (CALIBR) Menu

4.1

Front Panel Calibration

To calibrate the indicator using the front panel, do the following:

1. Place the indicator in setup mode (display reads

CONFIG

) and remove all weight from the scale platform.

If your test weights require hooks or chains, place the hooks or chains on the scale for zero calibration.

2. Press until the display reads

CALIBR

(see Figure 4-1). Press to go to zero calibration (

WZERO

).

3. With

WZERO

displayed, press

to calibrate zero. The indicator displays

*CAL*

while calibration is in

progress. When complete, the A/D count for the zero calibration is displayed. Press

again to save the

zero calibration value and go to the next prompt (

WVAL

).

4. With

WVAL

displayed, place test weights on the scale and press

to show the test weight value. Use the

numeric keypad to enter the actual test weight, then press

to save the value and go to span calibration

(

WSPAN

).

5. With

WSPAN

displayed, press

to calibrate span. The indicator displays

*CAL*

while calibration is in

progress. When complete, the A/D count for the span calibration is displayed. Press

again to save the

span calibration value and go to the next prompt (

REZERO

).

6. The rezero function is used to remove a calibration offset when hooks or chains are used to hang the test

weights.

If no other apparatus was used to hang the test weights during calibration, remove the test weights and
press to return to the CALIBR menu.

If hooks or chains were used during calibration, remove these and the test weights from the scale. With
all weight removed, press

to rezero the scale. This function adjusts the zero and span calibration

values. The indicator displays

*CAL*

while the zero and span calibrations are adjusted. When

complete, the adjusted A/D count for the zero calibration is displayed. Press

to enter the value, then

press to return to the CALIBR menu.

7. Press until the display reads

CONFIG

, then press to exit setup mode.

4.2

EDP Command Calibration

To calibrate the indicator using EDP commands, the indicator EDP port must be connected to a terminal or
personal computer. See Section 2.3.3 for EDP port pin assignments; see Section 5.0 for more information about
using EDP commands.

Once the indicator is connected to the sending device, do the following:

1. Place the indicator in setup mode (display reads

CONFIG

) and remove all weight from the scale platform.

If your test weights require hooks or chains, place the hooks or chains on the scale for zero calibration.