Configuration

23

3.1.2

Serial Command Configuration

The serial command set can be used to configure the

920i

indicator using either a personal computer,

terminal, or remote keyboard. Like

iRev 4

, serial

command configuration sends commands to the
indicator serial port; unlike

iRev 4

, serial commands

can be sent using any external device capable of
sending ASCII characters over a serial connection.

Serial commands duplicate the functions available
using the indicator front panel and provide some
functions not otherwise available. Serial commands
can be used to simulate pressing front panel keys, to
configure the indicator, or to dump lists of parameter
settings. See Section 9.0 on page 89 for more
information about using the serial command set.

3.1.3

Front Panel Configuration

Use the CONFIG submenu under the SCALES menu
to configure A/D scales. For example, in an indicator
with a single-channel A/D card installed in Slot 1, the
Scale Configuration display will show the A/D listed
(

Slot 1 Channel 1

) under the

AVAILABLE A/D’s

column.

Use the

left

navigation key to select the A/D, then

press the center softkey,

Add

. The A/D is then moved

to the Associated A/D’s column. If no other A/D’s are
listed in the

AVAILABLE A/D’s

column, the center

softkey changes to

Done

, as shown in Figure 3-2.

Press

Done

to exit the Scale Configuration display.

See Section 10.4 on page 111 for information about
configuring serial scales.

Figure 3-2. Scale Configuration Display

3.1.4

Multi-Range and Multi-Interval Scales

The

920i

supports multi-range and multi-interval

scales of either two or three ranges or intervals.

Multi-range scales provide two or three ranges, each
extending from zero to the maximum capacity
specified for the range, that can specify different scale
intervals (graduations). The scale interval changes as
the applied weight increases but does not reset to
lower range intervals until the scale returns to zero.

Multi-interval scales divide the scale into two or three
partial weighing ranges, each with different scale
intervals. The scale interval changes with both
increasing and decreasing loads applied.

To configure a multi-range or multi-interval scale, use
the SPLIT parameter to select 2RNG or 3RNG (for
multi-range scales), or 2INTVL or 3INTVL (for
multi-interval scales). Selecting a SPLIT value other
than OFF allows specification of decimal point,
display divisions, and maximum capacity for each
range or interval.

The SPLIT parameter is used to enable multi-range or
multi-interval. The SPLIT parameter is in the
SCALES menu, see Figure 3-4, and Table 3-2. After
setting the SPLIT parameter, the

Format

menu

selection will change as shown in Figure 3-6, and
Table 3-4.

If using streaming with multi-range or multi-interval,
the stream must be set to Custom in

iRev 4

. The

Tokens for Secondary and Tertiary Units must be set
to L or K to match the Primary, refer to the Serial
Menu, Tokens Parameter, in Section 3.2.2 on page 37.
They can be set using

iRev 4

or through the front

panel.

In multi-range, each range has its own capacity and
display division, extending from zero. The scale
display division will increase at the entered range
capacities, either two or three ranges. Once the range
has increased to the next level, the display division
will remain in new range until the scale returns to
zero. The tare value can be taken in any range.

For example,

Range 1 is 0 - 3000 x 1 lb.
Range 2 is 0 - 10,000 x 5 lb.

In multi-interval, the scale has one capacity, which is
segmented into weighing intervals, either two or three
intervals, each with different display division sizes.
As the weight value exceeds an interval or set interval,
the display division will increase, as the weight falls
below an interval or set interval, the display division
will decrease. The tare can only be taken in the first
interval.

For example,

Range 1 is 0-30 x 0.01 lb.
Range 2 is 30 - 60 x 0.02 lbs.