4.4

Dynamic Weighing

Note: Before using any application, be sure the balance has been leveled and calibrated.

Use this application to weigh an unstable load, such as a moving animal. Two different start/reset modes can be
selected: Manual (start and stop via key press) and Automatic (start and stop automatically).

Dynamic Weighing

– Manual (default)

1. In the upper left portion of the home screen, select Dynamic Weighing

2. Place objects on the pan and press the Start button.

The Dynamic Weighing Home screen

Main Display Line

Reference Fields

Application Buttons

Application

Icon

3. The balance begins a countdown (averaging process). During the countdown, the information line

displays the time remaining.

If necessary, press Stop to quit.

4. When the countdown ends, the result line is displayed and held. Press Clear to clear the held weight and

return to the initial screen.

Note: Clear the pan before beginning a new Dynamic weighing cycle.

4.4.1

Item Settings

To view or adjust the current settings

Touch the Item Settings button. The Settings screen

appears.

Averaging Time: Set the averaging time to a value
between 1 and 99 seconds. Default is 5 seconds.

Automatic Mode: When set to On, the cycle begins
when an object is placed on the pan, and the held value
is automatically reset when the object is removed from
the pan.

Sample Name: Assign a name to the sample.

Print settings:
Change printing settings. See section 7
for more information.

Note: The Save to USB button will only appear after a
USB flash drive has been connected to the balance.
See section 4.1.3 for more information.