9.4

Communication

9.4.1

Interface Commands

Commands listed in the following table will be acknowledged by the balance.
The balance will return “ES” for invalid commands.

TABLE 9-6. ADVENTURER INTERFACE COMMAND LIST

Command
Characters

Function

IP

Immediate Print of displayed weight (stable or unstable).

P

Print displayed weight (stable or unstable).

CP

Continuous Print.

SP

Print on Stability.

SLP

Auto Print stable non-zero displayed weight.

SLZP

Auto Print stable non-zero weight and stable zero reading.

xP

Interval Print x = Print Interval (1-3600 sec) 0P ends interval Print

0P

See above

H

Enter Print Header Lines

Z

Same as pressing Zero Key

T

Same as pressing Tare Key.

xT

Establish a preset Tare value in displayed unit. X = preset tare value. Sending 0T clears tare (if
allowed).

PT

Prints Tare weight stored in memory.

ON

Brings out of Standby

OFF

Goes to Standby.

C

Begin Span Calibration

IC

Begin internal Calibration, same as trigger from calibration menu.

AC

Abort Calibration. Attention: when LFT ON, the operation is not allowed.

PSN

Print Serial Number.

PV

Print terminal software version, base software version and LFT ON (if LFT is set ON).

x#

Set Counting APW (x) in grams. (must have APW stored)

P#

Print Counting application APW.

x%

Set Percent application reference weight (x) in grams. (must have reference weight stored)

P%

Print Percent application reference weight.

xS

0 = print unstable data, same as IP; 1 = print stable only

1)

, same as SP.

xRL

0 = disable response; 1 = enable response. This command only controls the “OK!” response.