8.

MAINTENANCE

8.1

Calibration

Periodically verify calibration by placing an accurate weight on the balance and viewing the result. If calibration is
required, refer to section 5.2 for instructions.

8.2

Cleaning

WARNING: Disconnect the Adventurer Balance from the power supply before cleaning.
Make sure that no liquid enters the interior of the balance.

Clean the Balance at regular intervals.

Housing surfaces may be cleaned with a lint-free cloth slightly dampened with water or a mild
cleaning agent.

Glass surfaces may be cleaned with a commercial glass cleaner. Please follow the steps below
on how to remove and install the sliding doors.

Attention: Do not use solvents, harsh chemicals, ammonia or abrasive cleaning agents.

Removing and reinstalling the glass doors for cleaning:

Step 1.
On the back of the balance, press
the pin and slide out the door.

Step 2.
After cleaning, slide the doors into
the slot while pressing the pin
mentioned in step 1.

Step 3.
Slide the doors into the slot until the
back stopper aligns with the other
door.

8.3

Troubleshooting

TABLE 8-1. TROUBLESHOOTING

Symptom / Display

Possible Cause

Remedy

Balance will not turn on No power to Balance

Verify connection and voltage

Poor accuracy

Improper calibration
Unstable environment

Perform calibration
Move balance to suitable location

Cannot calibrate

Calibration Menu locked
Approved Mode set to on

Unstable environment
Incorrect calibration masses

Turn Calibration menu lock off
Turn Approved Mode off

Move balance to suitable location
Use correct calibration masses

Cannot change menu
settings

Sub-menu locked

Approved Mode set to on

Unlock sub-menu
Turn

Approved Mode

off

Low Reference weight

Reference weight too small
The weight on the pan is too small to
define a valid reference weight.

Increase sample size

Invalid Piece Weight

Average piece weight is too small

Increase average piece weight

Operation Timeout

Weight reading is not stable

Move balance to suitable location

------

Busy (tare, zero, printing, waiting for a
stable weight)

Wait until completion

8.4

Service Information

If the troubleshooting section does not resolve your problem, contact an Authorized Ohaus Service Agent. Please
visit our website www.ohaus.com to locate the Ohaus office nearest you. An Ohaus Product Service Specialist will
be available to assist you.