ADVENTURER BALANCES

EN-31

5.2.6

Linearity Calibration

Linearity calibration uses three calibration points, one at zero load and the others at specified loads.
With no load on the scale, press Linearity Calibration to begin the process.
The balance captures the zero point, and then prompts for the next weight.
Continue to follow the instructions on the display until the calibration is completed.

To cancel at any time, press Cancel.

5.2.7

Calibration Test

Use Calibration Test to compare a known calibration weight against the stored span calibration data.

With no load on the scale, press Cal Test to begin the process.
The balance captures the zero point, then prompts for the span value.

The display shows status, followed by the difference between the current calibration weight and the
stored calibration data.


5.3 Balance Setup

Enter this sub-menu to customize Balance functionality.

5.3.1

Balance Set-up sub-menu

Language

User Settings

Filter Level

Auto Zero

Tracking

Auto Tare

Graduations

Date & Time

Approved Mode


Factory default settings are shown below in bold.

5.3.2

Language

Set the language displayed for menus and displayed
messages.


English
German
French
Spanish
Italian
Russian
Chinese
Japanese
Korean