Note

Trimming & Calibrating

11

Trimming procedures must match in order to proceed to weigh pad/section trimming in Step 10.

5. Pull excess cable out of the enclosure and tighten the cord grip assemblies with a wrench. To be watertight,

each cord grip must be tightened so the rubber sleeve begins to protrude from the hub.

6. Unused hubs must be properly plugged to prevent moisture entry.
7. Remove the desiccant from the plastic bag, and insert the desiccant bag into the junction box before

closing. Inspect the desiccant during normal service and change the desiccant as needed.

8. Replace the cover and torque the cover screws in an alternating pattern 15 in/labs to be certain the gasket is

compressed equally in all locations.

9. When the homerun cable is protected and in its final position, complete connection to the indicator. See the

indicator installation manual for wiring information.

10. Perform calibration using 25% of the Rice Lake AX3040 total capacity referring to the indicator manual to

determine the correct calibration procedure.