64

Counterpart User Manual

6.1.5

Parameter Setting Commands

Parameter setting commands allow you to display or
change the current value for a particular configuration
parameter (Tables 6-4 through 6-10).

Current configuration parameter settings can be
displayed in all modes using the following syntax:

command<ENTER>

Most parameter values can be changed in menu mode
only.

Use the following command syntax when changing
parameter values:

command=value<ENTER>

, where

value

is either a number or a parameter value. Use no spaces
before or after the equal (=) sign. If you type an
incorrect command, the display reads

??

.

For example, to set the motion band parameter to 5
divisions, type the following:

SC.MOTBAND=5<ENTER>

For parameters with selectable values, enter the
command and equal sign followed by a question
mark:

command=?<ENTER>

to see a list of those values.

The indicator must be in menu mode to use this
function.

Some parameters are valid only if other
parameters or parameter values are
specified. Restrictions for front-panel

configuration also apply to serial command configuration.

Note

Command

Description

Values

SC.GRADS#n

Graduations

1–100000

SC.ZTRKBND#n

Zero track band

0, 0–100

SC.ZRANGE#n

Zero range

1.900000, 0–100

SC.MOTBAND#n

Motion band

1, 0–100

SC.SSTIME#n

Standstill time

1–65535

SC.OVRLOAD#n

Overload

FS+2%, FS+1D, FS+9D, FS

SC.DIGFLTR1#n
SC.DIGFLTR2#n
SC.DIGFLTR3#n

Digital filtering

1, 2, 4, 8, 16, 32, 64, 128, 256

SC.DFSENS#n

Digital filter cutout sensitivity

2OUT, 4OUT, 8OUT, 16OUT, 32OUT, 64OUT,
128OUT

SC.DFTHRH#n

Digital filter cutout threshold

NONE, 2D, 5D, 10D, 20D, 50D, 100D, 200D,
250D

SC.THRESH#n

Zero threshold

09999999

SC.SMPRAT#n

Sample rate

30HZ, 60HZ, 120HZ, 240HZ, 480HZ, 960HZ

SC.PWRUPMD#n

Power up mode

GO, DELAY

SC.TAREFN#n

Tare function

BOTH, NOTARE, PBTARE, KEYED

SC.PRI.DECPNT#n

Primary units decimal position

8.888888, 88.88888, 888.8888, 8888.888,
88888.88, 888888.8, 8888888, 8888880,
8888800

SC.PRI.DSPDIV#n

Primary units display divisions

1D, 2D, 5D

SC.PRI.UNITS#n

Primary units

lb, kg, g, oz, OFF

SC.ACCUM#n

Accumulator enable

ON, OFF

SC.WZERO#n

Zero calibration

SC.WVAL#n

Test weight value

test_weight_value

SC.WSPAN#n

Span calibration

SC.WLIN.F1–
SC.WLIN.F5#n

Actual raw count value for linearization
points 1–5

0–16777215

Table 6-4. SCALES Serial Commands