8 test menu, Figure 3-14. test menu test menu, Table 3-10. test menu parameters – Rice Lake Counterpart Configurable Counting Indicator - User Manual - Version 2.1 User Manual

Page 52

Advertising
background image

46

Counterpart User Manual

3.8

Test Menu

Figure 3-14. Test Menu

TEST Menu

Parameter

Choices

Description

Level 2 submenus

A/D

AD1
AD2
EXCVAC

Gives details of current or live A/D counts as well as stored zero and span A/D
values.

Shows voltage levels for signal and excitation voltages.

DIG I/O

LOOP
DIO 1
DIO 2
DIO 3
DIO 4

Tests your digital I/O ports. If they are functioning, “PASS” is displayed. If they are
not functioning, “FAIL” is displayed.

Both inputs and outputs are active low. They go to a ground
state when active.

The I/O ports become activated when the test is
performed. Make sure any equipment is disconnected
prior to performing this test to avoid it being
inadvertently activated.

COMM

PORT1
PORT2

Performs a loopback test on the serial ports. If they are functioning, “PASS” is
displayed. If they are not functioning, “FAIL” is displayed.

RAM

TEST

Tests the unit’s memory. if it is functioning, “PASS” is displayed. If it is not
functioning, “FAIL” is displayed.

KEYPAD

TEST

Tests the unit’s individual keypad buttons by displaying the name of the key
pressed. If nothing is displayed, the key is not functioning.

Press the Menu key to exit the test.

Level 3 submenus - Under Level 2 A/D Feature

RAW

VAL

Displays the live current raw A/D count.

ZERO

VAL

Displays the captured A/D Zero calibration value.

SPAN

VAL

Displays the captured A/D Span calibration value.

mV

VAL

Displays the live current millivolt signal voltage.

EXCVDC

VAL

Displays the current excitation voltage.

LOOP

TEST

Performs a loop test on dig I/O cards.

Table 3-10. TEST Menu Parameters

A/D

DIG I/O

COMM

KEYPAD

RAM

LOOP

DIO 1

DIO 2

DIO 3

STATUS

DIO 4

STATUS

STATUS

STATUS

PORT1

PORT2

TEST

SPAN

mV

RAW

ZERO

VAL

VAL

VAL

VAL

EXCVDC

VAL

XXXXXXX

ID

DISPLAY

TEST

SETUP

AUDIT

CALIBR

WARNING...

DISCONNECT

DIGITAL I/O

BEFORE

RUNNING TEST

PASS

or

FAIL

PASS

or

FAIL

PASS

or

FAIL

PASS

or

FAIL

AD1

AD2

VAL

VAL

SETPTS

Note

Warning