0 appendix, 1 error messages, 2 using the xe and xeh edp commands – Rice Lake Counterpart Configurable Counting Indicator - User Manual - Version 2.1 User Manual

Page 84: Appendix, 1 error messages 9.2 using the xe edp command

Advertising
background image

78

Counterpart User Manual

9.0

Appendix

9.1

Error Messages

If an error code appears on the display, use the information in Table 9-1 as a troubleshooting guide. If you cannot
clear the error, call RLWS Service for assistance.

9.2

Using the XE and XEH EDP Commands

The XE and XEH EDP commands can be used to remotely query the

Counterpart

for the error conditions shown

on the front panel. The XE command returns a 5-digit number in the format:

xxxxx

where

xxxxx

contains a decimal representation of any existing error conditions as described in Table 9-2.

The XEH command returns a value in the format:

0xnnnnnnnn

where nnnnnnnn contains a hexadecimal representation of any existing conditions as described in Table 9-2.

If more than one error condition exists, the number returned is the sum of the values representing the error
conditions. For example, if the XE command returns the number 1040, this value represents the sum of an A/D
reference error (1024) and an A/D calibration checksum error (16).

Error Display

Description

Solution

Over range

• Check load cell wiring, including sense

jumpers.

• Check configuration, including number of

grads, channel selection, display divisions.

• Check calibration, including W ZERO and W

SPAN values.

• Under range can be caused by OIML setting

and weight less than -20 display divisions

• Check for scale binding or damage.
• Check for proper excitation voltage.
• Call RLWS Service for assistance.

Under range

A/D out of range

CHKERR - BATTERY BACK LOST Battery Error

Press ENTER after replacing the battery to restore
defaults to battery backed storage.

ERROR

Internal program error

Check configuration. Run XE command (see
Section 9.2 on page 78) to determine error type.
Call RLWS Service if unable to clear error by
cycling power or if error reoccurs.

HWFERR

Hardware error

Reboot the unit. Call RLWS Service if unable to
clear error by cycling power or if error reoccurs.

INVALID UNITS

ID uses unassigned units

Modify the ID to use a configured unit.

NOTARE

Tare is prevented

Change regulatory mode settings or the TAREFN
parameter.

OVERFL

Overflow error

Weight value too large to be displayed.

RANGE

GRADS > 100,000
WVAL > 100,000

Only shows up in Config mode.

EEPERR

EEPROM error

Call Rice Lake Weighing Systems (RLWS) for
service

VERSION UPDATED

Core has been updated or memory
has been corrupted.

Press the ENTER key. If the message persists, call
RLWS for service.

Table 9-1.

Counterpart Error Messages

- - - - - -

- - - - - -

- - - - - -

- - - - - -