6

CW90B/90XB Bench Scale Installation Manual

4.0

Troubleshooting

Table 4-1. Troubleshooting Guide

Symptom

Probable Cause

Remedy

No display

Power disconnected

Connect power

Cable cut or disconnected

Repair cable

Signal leads incorrectly wired at indicator

Connect according to indicator manual

Display stays at zero

Incorrect load cell cable connections

Connect according to manual

Faulty indicator

Service the indicator

Erratic weight display

Vibration near the scale

Remove the source of vibration, or adjust the
digital averaging of the indicator to minimize
the erratic display

Scale is not level

Level the scale

Water damage to the load cell or load cell
cable

Replace the load cell

Faulty indicator

Service the indicator

Loose load cell screws

Tighten the correct torque

Faulty load cell

Test and replace if necessary

Consistently low weights

Indicator is not properly adjusted to zero

Zero the indicator correctly

Scale deck cover is binding

Obtain adequate clearance

Overload stops set too high

Reset overload stops correctly

Indicator not calibrated for scale

Calibrate

Faulty load cell

Test and replace if necessary