0 calibration, Calibration – Rice Lake CW-90X Base User Manual

Page 9

Advertising
background image

Calibration

5

3.0

Calibration

Note

See your indicator manual for specific indicator calibration procedures.

It is recommended that the scale be “exercised” by loading it with test weights. Do this two or three times before
calibrating to make sure that the scale is seated. To calibrate the CW-90B/90XB scale, place test weights on the
scale platform equal to 70 - 100% of the scale’s capacity. If several weights are used, distribute them evenly around
the platform.